product design

Espectrofotómetro para MicroplacasEpoch

Detalles Técnicos (Inglés)

General

Detection modes Absorbance
Read methods Endpoint, kinetic, spectral scanning, well area scanning
Microplate types 6- to 384-well plates
Other labware supported Take 3 Micro-Volume Plates
Temperature control not available
Shaking not available
Software Gen5™ Data Analysis Software
Gen5 Secure for 21 CFR Part 11 compliance (option)
Automation BioStack™ and 3rd party automation compatible ('R' configuration)

Absorbance

Light source Xenon flash
Detector photodiode
Wavelength selection monochromator
Wavelength range 200 to 999 nm, in 1 nm increments
Monochromator bandwidth 5 nm
Dynamic range 0 - 4.0 OD
Resolution 0.0001 OD
Pathlength correction yes
Monochromator wavelength accuracy + 2 nm
Monochromator wavelength repeatability + 0.2 nm
OD accuracy 0 to 2 OD: +1% +0.010 OD
2 to 2.5 OD: +3% +0.010 OD
OD linearity 0 to 2 OD: +% +0.010 OD
2 to 2.5 OD: +3% +0.010 OD
OD repeatability 0 to 2 OD: +1% +0.005 OD
2 to 2.5 OD: +3% +0.005 OD
Reading speed (kinetic) 96 wells: 15 seconds
384 wells: 31 seconds

Physical Characteristics

Power 100 - 240 Volts AC. 50/60 Hz.
Dimensions 12" W x 12.5" D x 7.7" H
(30.5 cm x 31.8 cm x 19.6 cm)
Weight <15 lbs (6.8 kg)

Regulatory

Regulatory CE and TUV marked. ROHS compliant. In Vitro Diagnostic use models are available.